FDA Adverse Event Malfunction Summary report: N

ACCESS® 2 IMMUNOASSAY CLINICAL SYSTEM

MDR report key: 2030074 · Received March 25, 2011

Report

Report Number
2122870-2011-00810
Event Type
Malfunction
Date Received
March 25, 2011
Date of Event
February 23, 2011
Report Date
February 23, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K922823/A007
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

QC WAS WITHIN THE ESTABLISHED RANGES PRIOR TO AND ON THE DAY OF THE EVENT. A SYSTEM CHECK WAS PERFORMED ON (B)(6) 2011 AND MET THE SPECIFICATIONS. SERVICE WAS ON SITE ON (B)(4) 2011, AND THE FIELD SERVICE ENGINEER (FSE) NOTED A LARGE AMOUNT OF BUBBLES FORMING IN THE WASH PUMP UPON INSPECTION OF THE FLUIDICS MODULE. THE FSE LOOKED FOR FLOW RESTRICTIONS WITHIN THE WASH PUMP LINES AND FOUND NONE. THE FSE CLEANED AND REASSEMBLED THE WASH PUMP, BUT BUBBLE FORMATION RE-OCCURRED. THE FSE REPLACED THE INTERIOR WASH PUMP PARTS AND THE BUBBLE FORMATION CEASED. SYSTEM CHECK PASSED INSTRUMENT SPECIFICATIONS AFTER SERVICING THE INSTRUMENT, AND QC WAS WITHIN THE ESTABLISHED RANGES. A DEFINITIVE ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED TO DATE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER, INC. (BCI) IN REGARDS TO ERRONEOUSLY ELEVATED TROPONIN (ACCUTNI) RESULTS, ABOVE THE AMI CUTOFF AND WITHIN THE RISK STRATIFICATION RANGE, GENERATED BY ACCESS 2 IMMUNOASSAY SYSTEM FOR THREE PATIENTS' SAMPLES. THE RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. SUBSEQUENT TESTING PRODUCED LOWER RESULTS WITHIN THE RISK STRATIFICATION RANGE FOR EACH PATIENT. THE CUSTOMER DID NOT RECEIVE ANY REPORT OF PATIENT INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED TO THIS EVENT.