FDA Adverse Event
Injury
Summary report: N
HEARTMATE II LVAS
MDR report key: 2030031
·
Received March 25, 2011
Report
- Report Number
- 2030031
- Event Type
- Injury
- Date Received
- March 25, 2011
- Date of Event
- March 21, 2011
- Report Date
- June 7, 2011
- Manufacturer
- THORATEC CORPORATION
- Product Code
- DSQ
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
MAJOR PUMP UNIT(S) INVOLVED: BLOOD PUMP. ADDITIONAL TEXT: PUMP FAILURE FOR < 1 MINUTE DUE TO POSSIBLE GROUNDING ISSUES. SPECIFIC COMPONENT(S) INVOLVED: DRIVELINE MALFUNCTION. ADDITIONAL TEXT: POSSIBLE GROUNDING ISSUE. OTHER COMPONENT: CAUSATIVE OR CONTRIBUTING FACTOR: NO SPECIFIC CONTRIBUTING CAUSE IDENTIFIED. OTHER CAUSE: INTERVENTION(S): OTHER INTERVENTIONS, SPECIFY. OTHER INTERVENTION : REPLACEMENT OF THE PATIENT PM CABLE WITH A SPECIAL ONE. IMPLANT DEVICE TYPE: LVAD.
Description of Event or Problem · 1
MAJOR PUMP UNIT(S) INVOLVED: BLOOD PUMPSPECIFIC COMPONENT(S) INVOLVED: DRIVELINE MALFUNCTION
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTMATE II LVAS | LVAD | DSQ | THORATEC CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR |