FDA Adverse Event Malfunction Summary report: N

GATEWAY BALLOON CATHETER

MDR report key: 2030002 · Received March 25, 2011

Report

Report Number
2939204-2011-00156
Event Type
Malfunction
Date Received
March 25, 2011
Date of Event
March 20, 2011
Report Date
March 20, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
GBA
PMA / PMN Number
H050001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A DEVICE HISTORY RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. INSPECTION OF THE RETURNED DEVICE FOUND NO DAMAGE TO THE BALLOON OR THE DISTAL TIP OF THE DEVICE. A 0.014 INCH DEMONSTRATION GUIDEWIRE WAS INSERTED INTO THE DEVICE AND AN ATTEMPT WAS MADE TO INFLATE THE BALLOON WITH WATER. ATTEMPTS TO INFLATE THE DEVICE TO NOMINAL PRESSURE WERE UNSUCCESSFUL. THE STRAIN RELIEF WAS CAREFULLY REMOVED TO EXPOSE THE PROXIMAL END OF THE INNER SHAFT. MAGNIFICATION INSPECTION CONFIRMED THAT THE INNER SHAFT WAS APPROPRIATELY SEATED IN THE HUB AND ADHESIVE PRESENT IN THE ADHESIVE CAVITY. MAGNIFIED VIEWS OF THE INNER SHAFT THROUGH THE INFLATION PORT IN THE HUB REVEALED SMALL SLIT IN THE INNER SHAFT DIRECTLY ALIGNED WITH THE INFLATION PORT. THE DAMAGE TO THE INNER SHAFT MAY BE DUE TO PERFORATION WITH A NEEDLE OR GUIDE WIRE. NO TOOLS ARE INSERTED IN THE MANIFOLD INFLATION PORT DURING CATHETER ASSEMBLY. FURTHERMORE, ALL DEVICES ARE TESTED FOR BALLOON INFLATION/DEFLATION PERFORMANCE WITH VACUUM DECAY TESTING DURING MANUFACTURE. FROM THE INFORMATION PROVIDED AND THE INVESTIGATION IT WAS DETERMINED THAT HANDLING DAMAGE WAS THE MOST LIKELY CAUSE OF THE REPORTED EVENT.

Description of Event or Problem · 1

DURING PREPARATION IT WAS FOUND THAT THE BALLOON LEAKED. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

DURING PREPARATION IT WAS FOUND THAT THE BALLOON LEAKED. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GATEWAY BALLOON CATHETER CATHETER, BALLOON TYPE GBA BOSTON SCIENTIFIC - MAPLE GROVE M0032072409200 13666888

Patients

Seq Age Sex Outcome Treatment
1 72 YR