FDA Adverse Event
Malfunction
Summary report: N
ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM
MDR report key: 20297906
·
Received September 24, 2024
Report
- Report Number
- 3006575795-2024-00830
- Event Type
- Malfunction
- Date Received
- September 24, 2024
- Date of Event
- August 30, 2024
- Report Date
- September 24, 2024
- Manufacturer
- ZYNO MEDICAL, LLC
- Product Code
- FRN
- UDI-DI
- 00814371020037
- PMA / PMN Number
- K130690
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THERE ARE NO PREVIOUS COMPLAINTS ON THE DEVICE. TESTING RESULTS WERE REVIEWED. IT WAS DISCOVERED THAT THERE WERE DISCREPANCIES IN THE TEST RECORDS. NCR #2024-018 HAD BEEN INITIATED TO ADDRESS THE DISCREPANCIES. THIS PUMP UNDERWENT ROUTINE MAINTENANCE TESTING BY "INTUVIE" PROVIDER WHERE IT WAS DETECTED THAT THE PRESSURE WAS SET AT 30 PSI FAILED AT 42. THE RECALIBRATION CONFIRMED TO HAVE SUCCESSFULLY ADDRESSED THE ISSUE. A CAPA HAS BEEN OPENED IN ORDER TO FULLY INVESTIGATE AND ADDRESS THE ROOT CAUSE OF THE REPORTED EVENT. REFERENCE TO COMPLAINT #(B)(4).
Description of Event or Problem · 0
ON 08/30/2024, ZNYO MEDICAL RECEIVED A REPORT THAT DURING THE PREVENTIVE MAINTENANCE (PM), THAT WHEN THE PRESSURE WAS SET AT 30 PSI, IT FAILED AT 42. NO REPORT PATIENT WAS INVOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1098579 | ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM | LARGE VOLUME INFUSION PUMP | FRN | ZYNO MEDICAL, LLC | Z800WF | 191105292 | 00814371020037 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Prefer Not To Disclose |