FDA Adverse Event Malfunction Summary report: N

MERGE HEMO

MDR report key: 20297252 · Received September 24, 2024

Report

Report Number
2183926-2024-00004
Event Type
Malfunction
Date Received
September 24, 2024
Date of Event
August 26, 2024
Report Date
August 26, 2024
Manufacturer
MERGE HEALTHCARE
Product Code
DQK
UDI-DI
00842000100966
PMA / PMN Number
K152864
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

UPDATED THE FOLLOWING FIELDS: MEDSUN # (B)(4) IN H10, 510K NUMBER K152864 IN G4, TYPE OF REPORT FOLLOW UP 001 IN G6.

Additional Manufacturer Narrative · 0

MERGE HEALTHCARE HAS PERFORMED A COMPREHENSIVE INVESTIGATION AND RISK TO HEALTH ANALYSIS OF THE REPORTED ISSUE. THE HEALTH HAZARD EVALUATION CONCLUDED THAT THE EVENT IS UNLIKELY TO RESULT IN ANY ADVERSE HEALTH CONSEQUENCES. NO PATIENT HARM HAS OCCURRED AS A RESULT OF THE EVENTS REPORTED BY THE CUSTOMER FACILITY. MERGE HEALTHCARE WILL CONTINUE TO FOLLOW UP WITH FACILITY STAFF TO PREVENT FUTURE OCCURRENCES OF THE ISSUE. REVISED INFORMATION CONTAINED IN THIS SUPPLEMENTAL REPORT INCLUDE THE FOLLOWING: G6 - INDICATION THAT THIS IS FOLLOW-UP REPORT 002. H2-INDICATION OF ADDITIONAL INFORMATION. THIS IS THE FINAL REPORT FOR 2183926-2024-00004.

Description of Event or Problem · 0

ON 08/26/2024, MERGE HEALTHCARE RECEIVED A COPY OF A MEDSUN REPORT. A MERGE HEMO CUSTOMER REPORTED THAT THE CLIENT/DOCUMENTATION SIDE OF THE DEVICE SHUT DOWN DURING A ST SEGMENT ELEVATION MYOCARDIAL INFARCTION PROCEDURE. THE PHYSICIAN WAS ABLE TO PROCEED WITH THE PROCEDURE AND CARE FOR THE PATIENT BECAUSE MERGE HEMO CONTINUED TO DISPLAY AND UPDATE THE PATIENT'S VITAL SIGNS THROUGHOUT THE PROCEDURE. THE CUSTOMER RELOGGED INTO THE MERGE HEMO CLIENT SO THAT DOCUMENTING COULD CONTINUE. MERGE HEALTHCARE HAS INVESTIGATED THE ISSUE AND DETERMINED THAT THIS ISSUE OCCURRED DUE TO A SOFTWARE DEFECT. MERGE HEALTHCARE IS WORKING WITH THE CUSTOMER TO PROVIDE A SOFTWARE UPDATE. THERE HAVE BEEN NO REPORTS OF PATIENT INJURY OR HARM BECAUSE OF THIS ISSUE. REFERENCE: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2240153 MERGE HEMO COMPUTER, DIAGNOSTIC, PROGRAMMABLE DQK MERGE HEALTHCARE 10.4.1 00842000100966

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown