FDA Adverse Event Injury Summary report: N

EVOQUE TRICUSPID VALVE REPLACEMENT SYSTEM

MDR report key: 20295148 · Received September 24, 2024

Report

Report Number
2015691-2024-07339
Event Type
Injury
Date Received
September 24, 2024
Date of Event
September 7, 2024
Report Date
November 5, 2024
Manufacturer
EDWARDS LIFESCIENCES
Product Code
NPW
UDI-DI
00690103212846
PMA / PMN Number
P230013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE EVENT IS CAPTURED BY EDWARDS LIFESCIENCES UNDER COMPLAINT #: (B)(4). THE MANUFACTURER'S INVESTIGATION IS ONGOING AND A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Additional Manufacturer Narrative · 0

THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED/ADDED: B4, E1, G3, G6, H2, H6 AND H11. THE COMPLAINT FOR VALVE INTERACTS WITH CONDUCTION SYSTEM WAS CONFIRMED WITH OTHER EMPIRICAL EVIDENCE VIA INFORMATION PROVIDED BY EDWARDS CLINICAL SPECIALIST. THE REPORTED CONDUCTION DISTURBANCE DOES NOT ALLEGE A MALFUNCTION THAT COULD BE RELATED TO AN EDWARDS MANUFACTURING DEFICIENCY. THE EVENT DOES NOT ALLEGE A LABELING ISSUE OR DEVICE RELATED INFECTION; THEREFORE, NO DHR, LOT HISTORY REVIEW, OR MANUFACTURING MITIGATIONS REVIEW ARE REQUIRED. SINCE NO MANUFACTURING NON-CONFORMANCES WERE FOUND AND NO LABELING, TRAINING, OR IFU DEFICIENCIES WERE IDENTIFIED, NO PREVENTATIVE OR CORRECTIVE ACTIONS WERE REQUIRED.

Description of Event or Problem · 0

EDWARDS RECEIVED NOTIFICATION OF A 52MM EVOQUE VALVE PROCEDURE IN TRICUSPID POSITION. THE FINAL VALVE DEPLOYMENT POSITION WAS 0-5MM UNDER P/A AND S/L. ON POST-OPERATIVE DAY (POD) 2 THE PATIENT DEVELOPED AV-BLOCK, THIRD DEGREE AND ON POD 4 THE PATIENT HAD A PERMANENT PACEMAKER IMPLANTED. THE PATIENT IS IN STABLE CONDITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2017199 EVOQUE TRICUSPID VALVE REPLACEMENT SYSTEM PERCUTANEOUSLY DELIVERED PROSTHESES AND TRICUSPID VALVES NPW EDWARDS LIFESCIENCES 9850EV52 00690103212846

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention