FDA Adverse Event Injury Summary report: N

DAVINCI XI

MDR report key: 20292707 · Received September 24, 2024

Report

Report Number
2955842-2024-19491
Event Type
Injury
Date Received
September 24, 2024
Date of Event
August 27, 2024
Report Date
August 27, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110744
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT HAS BEEN RETURNED TO INTUITIVE SURGICAL, INC. (ISI) FOR EVALUATION. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) PERFORMED A SITE VISIT TO FURTHER INVESTIGATE THE CUSTOMER REPORTED ISSUE. THE FSE WAS UNABLE TO REPRODUCE THE CUSTOMER REPORTED FAILURE MODE. HOWEVER, THE FSE IDENTIFIED THAT THE LEFT MASTER TOOL MANIPULATOR (MTM) WAS NOT BE SEATED CORRECTLY. THE MTM WAS RECALIBRATED; THERE WAS NO PART REPLACEMENT NEEDED. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. DUE TO INSUFFICIENT PROCEDURAL INFORMATION, FURTHER SYSTEM/INSTRUMENT LOG INVESTIGATION CANNOT BE PERFORMED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED THYMECTOMY PROCEDURE, THE LEFT MASTER TOOL MANIPULATOR (MTM) WAS STIFFER THAN THE RIGHT MTM. THE SURGEON COULD NOT WORK WITH THAT AND SWAPPED THE SURGEON SIDE CONSOLE (SSC). THE SURGERY WAS FINISHED SUCCESSFULLY. HOWEVER, DURING FOLLOW UP WITH THE INITIAL REPORTER, IT WAS LEARNED THAT DURING THE MTM ISSUE, THE "LEFT ARM" SUDDENLY WENT UP UNCONTROLLABLY WHICH LED TO A BLEED THAT WAS LATER MANAGED AND THERE WAS A 30-45 MINUTE DELAY. THE FOLLOWING INFORMATION IS UNKNOWN: THE ESTIMATED BLOOD LOSS VOLUME ASSOCIATED WITH THE COMPLICATION, WHAT SPECIFIC MEDICAL INTERVENTION WAS RENDERED DUE TO THE BLEED, AND WHAT SURGICAL TASK THE SURGEON WAS ATTEMPTING TO PERFORM WHEN THE EVENT OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2017037 DAVINCI XI SURGEON SIDE CONSOLE, SMART PEDALS NAY INTUITIVE SURGICAL, INC 380677-04 N/A 00886874110744

Patients

Seq Age Sex Outcome Treatment
1 NA Prefer Not To Disclose Required Intervention DA VINCI INSTRUMENTS AND ACCESSORIES