FDA Adverse Event Death Summary report: N

LTV 1150 VENTILATOR

MDR report key: 20292073 · Received September 24, 2024

Report

Report Number
3013421741-2024-00440
Event Type
Death
Date Received
September 24, 2024
Date of Event
August 12, 2024
Report Date
September 24, 2024
Manufacturer
VYAIRE MEDICAL INC
Product Code
CBK
UDI-DI
00845873002726
PMA / PMN Number
K051767
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

VYAIRE MEDICAL FILE IDENTIFICATION: (B)(4). H3: ROOT CAUSE FINDINGS: THE DEVICE EVENT LOG WAS REVIEWED AND THE INCIDENT WAS CONFIRMED DURING THE REVIEW. ON THE DATE OF THE INCIDENT IT WAS CONFIRMED THAT THE DEVICE WAS ALARMING AUDIBLY AND VISUALLY FOR MULTIPLE ALARMS FOR A TOTAL OF 18 MINUTES UNTIL THE DEVICE WAS MANUALLY POWERED OFF. THE DEVICE WAS TESTED AND SET TO VENTILATE WITH THE SAME PATIENT SETTINGS FOR A TOTAL OF 71 HOURS AND CONFIRMED TO BE FUNCTIONING AS INTENDED WITHOUT ANY FAULTS.

Description of Event or Problem · 0

IT WAS REPORTED TO VYAIRE MEDICAL THAT A PEDIATRIC PATIENT WAS ON THE LTV 1150 DEVICE VIA A TRACHEOSTOMY TUBE. ON THE MORNING OF (B)(6), THE DAY-SHIFT NURSE ARRIVED TO FIND THE TRACHEOTOMY HAD BECOME DISLODGED AT SOME POINT DURING THE EVENING AND WAS UNABLE TO GET A PULSE READING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1139112 LTV 1150 VENTILATOR VENTILATOR, CONTINUOUS, FACILITY USE CBK VYAIRE MEDICAL INC 18984-001 00845873002726

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Death