FDA Adverse Event Injury Summary report: N

LOGIC TIBIA PS MOD INSRT SZ 2 15MM

MDR report key: 20291311 · Received September 24, 2024

Report

Report Number
1038671-2024-03674
Event Type
Injury
Date Received
September 24, 2024
Date of Event
November 2, 2022
Report Date
May 13, 2025
Manufacturer
EXACTECH, INC.
Product Code
JWH
UDI-DI
10885862001757
PMA / PMN Number
K033883
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

D10: CONCOMITANT DEVICES: (3943095) 200-03-35 - ONE PEG PATELLA 35MM; (4059307) 02-012-45-2020 - LGC TIBIAL FIT TRAY CEM SZ 2F / 2T; (4075084) 02-010-01-0220 - LOGIC FEMORAL PS CEM LEFT SZ 2. THE PRODUCT ASSOCIATED WITH THE REPORTED EVENT IS WITHIN THE SCOPE OF RECALL Z-0021-2022: HOWEVER, THERE IS INSUFFICIENT INFORMATION TO EVALUATE WHETHER THE SUBJECT ISSUE OF THE RECALL WAS A CAUSE OR CONTRIBUTOR TO THE REPORTED EVENT. THE DEVICE WAS NOT RETURNED FOR EVALUATION AND NO MEDICAL OR OTHER RECORDS CONTAINING TREATMENT INFORMATION OR PATIENT INFORMATION HAVE BEEN RECEIVED; THEREFORE, THE REPORTED EVENT CANNOT BE CONFIRMED, NOR CAN THE CIRCUMSTANCES OR POTENTIAL CAUSES OR CONTRIBUTORS TO THE ALLEGED EVENT BE EVALUATED. SHOULD ADDITIONAL, MATERIAL INFORMATION BECOME AVAILABLE THAT PERMITS MORE ANALYSIS OR CONCLUSIONS, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED: H6. MDR SECTION CODES UPDATED/CORRECTED: A, B, E. THE REASON FOR THE REVISION REPORTED CANNOT BE CONFIRMED FROM THE INFORMATION PROVIDED BUT MAY BE THE RESULT OF PROSTHESIS WEAR AND LOSS OF RANGE OF MOTION OR DUE TO INCLUSION OF THE POLYETHYLENE IN THE PACKAGING RECALL. ADDITIONALLY, A CONTRIBUTING FACTOR TO THE REPORTED WEAR MAY HAVE BEEN THE RESULT OF BEING PACKAGED IN A NON-CONFORMING VACUUM BAG FOR MORE THAN FIVE YEARS. POTENTIAL CONTRIBUTIONS OF USER AND PATIENT-RELATED CONSIDERATIONS TO THE EVENT COULD NOT BE ASSESSED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND IMAGES, RADIOGRAPHS AND RELEVANT CLINICAL INFORMATION WERE NOT PROVIDED. IF ANY FURTHER INFORMATION IS OBTAINED THAT WOULD CHANGE OR ALTER ANY INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 0

IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 86 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, JOINT PAIN, JOINT SWELLING AND STIFFNESS, LIMITED ROM, LOSS OF MOBILITY, REVISION SURGERY, POLYETHYLENE WEAR, CONSTRAINED LINER REVISION.. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1291149 LOGIC TIBIA PS MOD INSRT SZ 2 15MM PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH EXACTECH, INC. 10885862001757

Patients

Seq Age Sex Outcome Treatment
1 NA Female Hospitalization| R SEE H11