FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 20291031 · Received September 24, 2024

Report

Report Number
2245578-2024-00176
Event Type
Malfunction
Date Received
September 24, 2024
Date of Event
September 15, 2024
Report Date
October 29, 2024
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
MMI
PMA / PMN Number
K031739
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 22-OCT-2024. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AN (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED FOR CTNI CARTRIDGE LOT S24133.

Description of Event or Problem · 0

ON 17-SEP-2024, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DESCREPANT RESULTS ON A PATIENT. THERE WAS NO PATIENT INFORMATION. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD DATE TESTED RESULTS I-STAT 15-SEP-2024 08:47 0.24 NG/ML I-STAT 15-SEP-2024 11:12 0,02 NG/ML I-STAT 15-SEP-2024 11:26 0.19 NG/ML I-STAT 15-SEP-2024 11:37 0.02 NG/ML I-STAT 15-SEP-2024 11:37 0.02 NG/ML I-STAT 15-SEP-2024 12:22 0.02 NG/ML . AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THERE IS NO EVIDENCE THE PATIENT SUFFERED AN MI. THE INVESTIGATION IS UNDERWAY. PER I-STAT SYSTEM MANUAL: ART: 715595-00V REPORTABLE RANGE: 0.00 - 50.00 REFERENCE RANGE: 0.00 - 0.03 (REPRESENTS THE 0 TO 97.5% RANGE OF RESULTS) REFERENCE RANGE: 0.00 - 0.08 (REPRESENTS THE 0 TO 99% RANGE OF RESULTS).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1290139 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE INC. NA S24133

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown