NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2011-00192
- Event Type
- Malfunction
- Date Received
- March 24, 2011
- Date of Event
- September 29, 2011
- Report Date
- February 24, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT DEVICE HAS RECENTLY BEEN RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
(B)(4). METHOD: THE RETURNED NEOPUFF WAS VISUALLY INSPECTED FOR DAMAGE. RESULTS: TWO CRACKS WERE FOUND ON THE FRONT FASCIA OF THE NEOPUFF. THE UPPER AND LOWER END CAPS WERE ALSO FOUND TO BE CRACKED. THE TUBE THAT CONNECTS THE CENTER MANIFOLD PORT TO THE MANOMETER WAS CUT AT THE PORT END. THE CUT IS NOT SMOOTH AND THE TUBE AROND THE CUT WAS COMPRESSED. A 360-DEGREE CRACK WAS FOUND ALONG THE CIRCUMFERENCE OF THE MANIFOLD. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 070219. CONCLUSION: THE CUT IN THE TUBE APPEARS TO HAVE BEEN MADE INTENTIONALLY USING A TOOL AS THE TUBE IS COMPRESSED AROUND THE CUT. IT IS LIKELY THAT THE MANIFOLD CRACKED DUE TO STRESS WHEN THE TUBING WAS CUT. THE CUT WOULD HAVE CAUSED THE FAULT REPORTED BY THE CUSTOMER HOWEVER IT IS NOT KNOWN WHY THE TUBE WAS CUT. THE NEOPUFF USER INSTRUCTIONS REQUIRE THAT THE NEOPUFF BE TESTED BY QUALIFIED PERSONNEL OR AN AUTHORIZED FISHER & PAYKEL HEALTHCARE REPRESENTATIVE PRIOR TO EACH USE TO ENSURE FUNCTIONALITY. THE SET-UP PROCEDURE STATES THE USER MUST CHECK THE SETTINGS BY TESTING THE PRESSURE OUTPUT FROM THE NEOPUFF AT THE DESIRED FLOW RATE "PRIOR TO EVERY USE OF THE NEOPUFF TO ENSURE THAT THE DEVICE IS FUNCTIONING CORRECTLY".
A HOSPITAL IN (B)(6) REPORTED THAT AN RD900 NEOPUFF INFANT RESUSCITATOR DID NOT PASS THE PERFORMANCE CHECK. THE PRESSURE OUTPUT AT 6 LPM FLOW WAS TOO LOW. THE PERFORMANCE CHECK WAS CARRIED OUT PRIOR TO PATIENT USE.
A HOSPITAL IN (B)(6) REPORTED THAT AN RD900 NEOPUFF INFANT RESUSCITATOR DID NOT PASS THE PERFORMANCE CHECK. THE PRESSURE OUTPUT AT 6LPM FLOW WAS TOO LOW. THE PERFORMANCE CHECK WAS CARRIED OUT PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LIMITED | RD900 | 070219 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |