FDA Adverse Event Malfunction Summary report: N

O-ARM O2 IMAGING SYSTEM

MDR report key: 20277483 · Received September 23, 2024

Report

Report Number
3006544299-2024-00710
Event Type
Malfunction
Date Received
September 23, 2024
Date of Event
September 15, 2024
Report Date
January 20, 2025
Manufacturer
SANMINA -SCI SYSTEMS
Product Code
OWB
UDI-DI
00763000616564
PMA / PMN Number
K200074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

UPDATED B5. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H2-3) THE TANK KIT WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. AFTER FUNCTIONAL TESTING AND VISUAL/PHYSICAL EXAMINATION, UNABLE TO CONFIRM REPORTED COMPLAINT "UNABLE TO PERFORM SPIN." HV TANK PASSED BENCH TEST. THE RESISTANCE BETWEEN POINTS P1 AND P3, P1 AND P2, P3 TO P2, J1M TO J1A, J1R TO J1A , J1B AND J1A, J1C AND J1A, J1C TO J1A, J1F TO J1G, C-S AND C-L ON THE HV TANK PASSED. THE RESULTS OF THE ANALYSIS CONCLUDED THAT NO FAILURE WAS FOUND. CODES: B01, C19, D14 CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: BI71000469, SERIAL/LOT #: T232077 REV. E: MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H2: ADDITIONAL INFORMATION IN B5. B3 UPDATED. H3: THE SYSTEM WAS SERVICED IN THE FIELD. IT WAS FOUND THAT THERE WAS NO X-RAY WITH 2D OR 3D. E40 AND E41 ERRORS OCCURRED. THE TANK KIT WAS REPLACED AND THE SYSTEM PERFORMED AS INTENDED. H6: CODES FDM B01, FDR C09, AND FDC D02 ARE APPLICABLE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

NO PARTS HAVE BEEN RECEIVED BY THE MANUFACTURER FOR EVALUATION. CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: B I71000901; PRODUCT ID: BI71000469; PRODUCT ID: BI71000469, SERIAL/LOT; MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

ADDITIONAL INFORMATION WAS RECEIVED. IT WAS A SPINAL PROCEDURE.

Description of Event or Problem · 0

ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT THE SPIN DID NOT GET INTERRUPTED. THERE WAS NO IMPACT ON PATIENT OUTCOME. THERE WAS A 20-MINUTE SURGICAL DELAY.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING AN IMAGING SYSTEM BEING USED DURING A PROCEDURE. IT WAS REPORTED THAT SITE WHEN TRYING TO PERFORM A SPIN, THE SYSTEM INDICATED IT WAS PERFORMING THE SPIN BUT NEVER COMPLETED ONE. SITE TRIED REBOOTING SYSTEM WITH NO RESOLUTION. SITE CONTINUED WITH THE USE OF THE OTHER IMAGING SYSTEM. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1487548 O-ARM O2 IMAGING SYSTEM INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM OWB SANMINA -SCI SYSTEMS BI70002000 00763000616564

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown "SEE H11...."