FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 2026715 · Received March 18, 2011

Report

Report Number
2531779-2011-01723
Event Type
Malfunction
Date Received
March 18, 2011
Report Date
February 18, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K032257
Removal / Correction Number
2531779-02/25/11-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. THERE IS NO INVESTIGATION NECESSARY. CARTRIDGES WITH LOT #B201581 WERE CONFIRMED TO BE DEFECTIVE. WHEN TESTED, FLUID WAS OBSERVED LEAKING FROM THE PLUNGER END OF THE CARTRIDGE.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT FIVE CARTRIDGES FROM THE SAME SHIPMENT, BUT DIFFERENT BOXES HAVE LEAKED WHERE THE PLUNGER MEETS THE BARREL OF THE CARTRIDGE. HE OBSERVED THAT INSULIN COLLECTED INSIDE THE PLUNGER AND OUTSIDE THE BARREL. THE PATIENT STATED THAT HE DOES NOT REMOVE THE PLUNGERS FROM THE CARTRIDGES; HE NEVER RE-USES CARTRIDGES; HE FILLS THEM TO 200 UNITS, BUT NEVER OVERFILLS THE CARTRIDGE. HE NOTED THAT THE CARTRIDGE COMPARTMENT WAS WET, BUT THERE WAS NO VISIBLE INSULIN UNDER THE DISPLAY LENS. THE PATIENT SUSPENDED INSULIN PUMP THERAPY AND USED AN ALTERNATE METHOD OF INSULIN DELIVERY UNTIL HE RECEIVED A NEW BOX OF CARTRIDGES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP CARTRIDGE LZG ANIMAS CORP. IR 1200/1250/2020/OTP B201581

Patients

Seq Age Sex Outcome Treatment
1