FDA Adverse Event Injury Summary report: N

IMPLANTABLE COLLAMER LENS (ICL)

MDR report key: 20264219 · Received September 20, 2024

Report

Report Number
2023826-2024-04396
Event Type
Injury
Date Received
September 20, 2024
Date of Event
July 17, 2024
Report Date
August 25, 2024
Manufacturer
STAAR SURGICAL
Product Code
QCB
UDI-DI
00840311322695
PMA / PMN Number
P030016
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: G4- PREMARKET IDENTIFICATION: DEVICE BLA: P030016 TO PMA/510(K): P030016, INITIALLY REPORTED PMA/510(K) NUMBER IN THE WRONG FIELD. CLAIM # (B)(4).

Additional Manufacturer Narrative · 0

MANUFACTURER NARRATIVE - SECONDARY SURGICAL INTERVENTION TO REMOVE/REPLACE/REPOSITION THE LENS IS IDENTIFIED IN THE LABELING AS A KNOWN POTENTIAL ADVERSE EVENT FOLLOWING ICL IMPLANTATION. (B)(4)

Description of Event or Problem · 0

A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED EXCESSIVE VAULT. IN A SEPARATE VISIT THE LENS WAS EXCHANGED FOR A REPLACEMENT LENS OF SHORTER LENGTH AND THE PROBLEM RESOLVED. CAUSE OF EVENT WAS REPORTED AS UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252932 IMPLANTABLE COLLAMER LENS (ICL) PHAKIC TORIC INTRAOCULAR LENS QCB STAAR SURGICAL VTICMO12.6 N/A 00840311322695

Patients

Seq Age Sex Outcome Treatment
1 24 YR Female Required Intervention CARTRIDGE MODEL: SFC-45 W/FTP, LOT# UNK| INJECTOR MODEL: MSI-PF, LOT# UNK