FDA Adverse Event Malfunction Summary report: N

BD BBL¿ TRYPTICASE¿ SOY AGAR (SOYBEAN-CASEIN DIGEST AGAR) PLATE 100PK

MDR report key: 20263974 · Received September 19, 2024

Report

Report Number
1119779-2024-00680
Event Type
Malfunction
Date Received
September 19, 2024
Date of Event
August 27, 2024
Report Date
October 17, 2024
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
JSG
UDI-DI
10382902212833
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: DURING MANUFACTURING OF MATERIAL 221283, MEDIA IS FORMULATED AND SENT THROUGH A HIGH TEMPERATURE SHORT TIME STERILIZER TO REMOVE BIOBURDEN. THE PETRI DISHES ARE SUBJECTED TO UV RADIATION TO DECREASE BIOBURDEN. THE PETRI DISHES ARE FILLED IN A POSITIVE PRESSURE HEPA FILTERED ENVIRONMENT. THE FILLED PLATES ARE COOLED AND IMMEDIATELY WRAPPED INTO SLEEVES TO DECREASE THE INTRODUCTION OF MICROBES. SLEEVES ARE THEN PACKAGED INTO CARTONS AND THEN TRANSFERRED TO A REFRIGERATED TRUCK (2 TO 8 DEGREES C) FOR SHIPMENT TO THE DISTRIBUTOR. BD DISTRIBUTORS ARE PROVIDED WITH THE STORAGE GUIDELINES FOR THE SHIPPING AND HANDLING OF BD MEDIA OF 2 TO 8 DEGREES C IN A DARK PLACE. THE COMPLAINT HISTORY WAS REVIEWED, AND NO OTHER COMPLAINTS HAVE BEEN TAKEN ON BATCH 4135030. NO RETENTION SAMPLES FOR THIS BATCH WERE AVAILABLE FOR INVESTIGATION. NO RETURN SAMPLES OR PHOTOS WERE RECEIVED FOR THE INVESTIGATION. THIS COMPLAINT CANNOT BE CONFIRMED. NO COMPLAINT TRENDS HAVE BEEN IDENTIFIED ON THIS PRODUCT FOR THIS DEFECT. BD WILL CONTINUE TO TREND COMPLAINTS FOR THIS DEFECT.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED PRIOR TO USING BD BBL¿ TRYPTICASE¿ SOY AGAR (SOYBEAN-CASEIN DIGEST AGAR) PLATE 100PK THAT ONE (1) SLEEVE OF MEDIA WAS MISSING THE PRODUCT LABEL. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED PRIOR TO USING BD BBL¿ TRYPTICASE¿ SOY AGAR (SOYBEAN-CASEIN DIGEST AGAR) PLATE 100PK THAT ONE (1) SLEEVE OF MEDIA WAS MISSING THE PRODUCT LABEL. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
226807 BD BBL¿ TRYPTICASE¿ SOY AGAR (SOYBEAN-CASEIN DIGEST AGAR) PLATE 100PK CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL JSG BECTON DICKINSON & CO. (SPARKS) 4135030 10382902212833

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown