FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG

MDR report key: 20255916 · Received September 19, 2024

Report

Report Number
1917413-2024-00876
Event Type
Malfunction
Date Received
September 19, 2024
Date of Event
August 14, 2024
Report Date
August 30, 2024
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
PJE
UDI-DI
30382903627951
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD RECEIVED 3 SAMPLES AND 1 PHOTO FOR INVESTIGATION. VISUAL INSPECTION OF THE SAMPLES CONFIRMED THE TUBE LABEL IS MISSING FROM 2 OF THE 3 SAMPLE TUBES. VISUAL EVALUATION OF THE PHOTO WAS PERFORMED AND REVEALED A MISSING LABEL ON THE TUBE. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE MISSING LABEL. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED PRIOR TO USE OF BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG, A TUBE LABEL WAS MISSING ON ONE TUBE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1488177 BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 9.0 MG BLOOD/PLASMA COLLECTION DEVICE FOR DNA TESTING PJE BECTON, DICKINSON & CO. (BROKEN BOW) 4073675 30382903627951

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown