FDA Adverse Event Injury Summary report: N

SUBLIME¿ RADIAL ACCESS 018 RX PTA DILATATION CATHETER

MDR report key: 20251878 · Received September 18, 2024

Report

Report Number
3005994106-2024-00014
Event Type
Injury
Date Received
September 18, 2024
Date of Event
August 16, 2024
Report Date
September 18, 2024
Manufacturer
CREAGH MEDICAL
Product Code
LIT
UDI-DI
05391525786360
PMA / PMN Number
K211044
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3: ONLY ONE SEGMENT WAS RETURNED. THIS SEGMENT INCLUDED THE CATHETER HUB TO RAPID EXCHANGE (RX) PORT. THE DISTAL PORTION OF THE RETURNED SECTION SHOWED OUTER CATHETER ELONGATION OF APPROXIMATELY 5CM PAST THE SKIVE.

Description of Event or Problem · 0

THE BALLOON CATHETER WAS USED FOR A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE. RADIAL ANGIOPLASTY WAS PERFORMED IN ORDER TO AVOID AN OPEN PROCEDURE. THE SITE REPORTED THE BALLOON BURST AFTER INFLATION TO APPROXIMATELY 10ATMS. THE PHYSICIAN WAS USING THE "KISSING BALLOON TECHNIQUE" FOR BIFURCATIONS. DURING REMOVAL, THE BALLOON CATHETER SHAFT SEPARATED FROM THE MAIN BODY NEAR THE RAPID EXCHANGE PORT DURING REMOVAL FROM A 6FR SHEATH. THE THOUGHT WAS THAT THE SITE DID NOT ALLOW FULL DEFLATION OF THE BALLOONS PRIOR TO PULLING BACK INTO THE SHEATH. THE PHYSICIAN HAD TO PERFORM A VESSEL CUTDOWN TO RETRIEVE THE SEPARATED SEGMENTS AND FINISH WITH ENDARTERECTOMY. THE SITE INDICATED THE BALLOONS SEPARATED IN THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA) AND PROXIMAL POPLITEAL ARTERY. MULTIPLE ATTEMPTS AT DEFLATION WERE MADE. THE WIRE WAS NOT REMOVED. THE BALLOON SEPARATED APPROXIMATELY 10-20 CM FROM THE BALLOON MARKER BAND. THE FURTHER INTERVENTION OF ENDARTERECTOMY WAS REQUIRED FOR THE CALCIFIED ILIAC ARTERY AND BALLOON SEGMENT REMOVAL. BOTH BALLOON SEGMENTS WERE INTACT UPON REMOVAL. THE ENTIRE PROCEDURE LASTED APPROXIMATELY SEVEN HOURS WITH A FOUR HOUR DELAY DUE TO THE BALLOON SEPARATION AND CUTDOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2464236 SUBLIME¿ RADIAL ACCESS 018 RX PTA DILATATION CATHETER Catheter, angioplasty, peripheral, transluminal LIT CREAGH MEDICAL SRA-DC18-50120 22301354 05391525786360

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention