FDA Adverse Event
Malfunction
Summary report: N
BIOPSYS MARKER CLIP WITH INTRODUCER
MDR report key: 202504
·
Received December 15, 1998
Report
- Report Number
- 1527736-1998-04179
- Event Type
- Malfunction
- Date Received
- December 15, 1998
- Date of Event
- December 3, 1998
- Report Date
- December 3, 1998
- Manufacturer
- BIOPSYS MEDICAL, INC.
- Product Code
- KNW
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
IT WAS REPORTED BY THE REP THAT THE DEVICE WAS USED DURING STEREOTACTIC BREAST BIOPSY. IT WAS REPORTED THE METAL TIP OF THE MICROMARK CLIP APPLIER DETACHED FROM THE APPLIER UPON REMOVAL FROM THE PATIENT. THE SURGEON STATED THAT SHE WAS UNABLE TO RETRACT THE TIP INTO THE CANNULA PRIOR TO REMOVAL BECAUSE IT WAS CAUGHT ON THE EDGE OF THE NOTCH IN THE CANNULA. THE TIP OF THE APPLIER WAS LEFT IN THE PATIENT. THERE WAS NO CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BIOPSYS MARKER CLIP WITH INTRODUCER | BIOPSYS | KNW | BIOPSYS MEDICAL, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |