DO NOT USE-NANOTITE TM CERTAIN¿ IMPLANT 4 X 15MM
Report
- Report Number
- 0001038806-2024-02092
- Event Type
- Injury
- Date Received
- September 18, 2024
- Date of Event
- January 15, 2024
- Report Date
- January 28, 2025
- Manufacturer
- BIOMET 3I
- Product Code
- DZE
- PMA / PMN Number
- K051461
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 951231. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. STERILIZATION RECORD WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 951231 FOR SIMILAR EVENTS AND ONE OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: PERI-IMPLANTITIS. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
ZIMVIE COMPLAINT NUMBER (B)(4). . D10: CONCOMITANT MEDICAL PRODUCT: NIOSS415, DO NOT USE-NANOTITE TM CERTAIN IMPLANT 4 X 15MM, LOT: 951231. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
THE DOCTOR REPORTS THAT THE DENTAL IMPLANTS LOCATED IN DENTAL POSITIONS NUMBER # 14 & #15 FAILED DUE TO PERI-IMPLANTITIS AND WERE REMOVED. THE PROCEDURE WAS NOT CONCLUDED BY PLACING ANOTHER IMPLANTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1343657 | DO NOT USE-NANOTITE TM CERTAIN¿ IMPLANT 4 X 15MM | DENTAL IMPLANT | DZE | BIOMET 3I | 951231 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Female | DENTAL IMPLANT SEE H10 NARRATIVE. |