FDA Adverse Event Malfunction Summary report: N

VIDEO-OPTIK "ENDOEYE 3D", 30°

MDR report key: 20246082 · Received September 18, 2024

Report

Report Number
9610773-2024-32603
Event Type
Malfunction
Date Received
September 18, 2024
Report Date
January 15, 2025
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
GCM
UDI-DI
04042761083492
PMA / PMN Number
K193026
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

E2/E3: NO INFORMATION AVAILABLE FOR THE OCCUPATION OF THE INITIAL REPORTER OR WHETHER THEY ARE A HEALTHCARE PROFESSIONAL. THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS NOT CONFIRMED. HOWEVER, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: POOR IMAGE QUALITY; COMPONENT FAILURE OF R-UNIT (CHARGED COUPLED DEVICE). BASED ON THE RESULTS OF THE INVESTIGATION, A ROOT CAUSE COULD NOT BE IDENTIFIED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

D2A COMMON DEVICE NAME: RIGID VIDEO LAPAROSCOPE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SUBJECT DEVICE SWITCHED BETWEEN 2D AND 3D ON ITS OWN, AND WHEN TOUCHED THE CABLE, IT SWITCHED AND MADE NOISE AT THE SAME TIME. WHEN LAST USED, IT SOMETIMES WENT DARK WHEN CONNECTING AND COULD NOT BE SEEN OUT OF RIGHT EYE. THE ISSUE WAS FOUND DURING THE MIDDLE OF AN UNSPECIFIED THERAPEUTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SUBJECT DEVICE SWITCHED BETWEEN 2D AND 3D ON ITS OWN, AND WHEN TOUCHED THE CABLE, IT SWITCHED AND MADE NOISE AT THE SAME TIME. WHEN LAST USED, IT SOMETIMES WENT DARK WHEN CONNECTING AND COULD NOT BE SEEN OUT OF RIGHT EYE. THE ISSUE WAS FOUND DURING THE MIDDLE OF AN UNSPECIFIED THERAPEUTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1343626 VIDEO-OPTIK "ENDOEYE 3D", 30° RIGID VIDEO LAPAROSCOPE GCM OLYMPUS WINTER & IBE GMBH WA50082A 04042761083492

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ELITE II