BABY CONTROL WALL MOUNT INFANT WARMER
Report
- Report Number
- 9611451-2011-00173
- Event Type
- Malfunction
- Date Received
- March 20, 2011
- Date of Event
- February 18, 2011
- Report Date
- February 22, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- FMT
- PMA / PMN Number
- K971461
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE AFFECTED IW980JEU INFANT WARMER UNITS ARE NOT EXPECTED TO BE RETURNED TO FPH FOR INSPECTION. WE ARE CURRENTLY IN THE PROCESS OF OBTAINING FURTHER INFORMATION FROM THE MEDICAL CENTER IN ORDER TO ASSIST US WITH THE ANALYSIS AND IDENTIFICATION OF A POSSIBLE ROOT CAUSE OF THE REPORTED FAULT. WE WILL PROVIDE A FOLLOW UP REPORT ONCE WE RECEIVE FURTHER INFORMATION AND HAVE COMPLETED OUR ANALYSIS.
(B)(4). SERIAL NUMBERS: (B)(4), DEVICE MANUFACTURE DATES: 08/25/2003; (B)(4), 08/25/2003; (B)(4), 09/19/2003; (B)(4), 09/19/2003; (B)(4), 09/19/2003; (B)(4), 10/21/2004. METHOD: THE HARNESSES OF THE SIX IW980JEU BABY CONTROL WALL MOUNT INFANT WARMERS WERE NOT RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH) IN (B)(4) FOR INSPECTION. OUR ANALYSIS IS BASED ON THE EVENT DESCRIPTIONS AND RESULTS OF PREVIOUS INVESTIGATIONS ON SIMILAR COMPLAINTS. RESULTS: THE MEDICAL CENTER IN (B)(6) REPORTED THAT THEIR SIX IW980JEU UNITS HAD DISCOLOURED UPPER AND LOWER HARNESS CONNECTORS. CONCLUSION: THE UPPER AND LOWER HARNESSES ARE USED TO CONNECT THE HEAD UNIT TO THE CONTROL UNIT OF THE INFANT WARMER. PREVIOUS INVESTIGATIONS ON SIMILAR COMPLAINTS REVEALED THAT THE DISCOLOURED HARNESS CONNECTORS OF THE INFANT WARMERS WERE MOST LIKELY DUE TO THE ARCING THAT OCCURRED BETWEEN TWO ELECTRICAL CONTACTS, RESULTING IN CONTACT FAILURE. THE ARCING WAS MOST LIKELY DUE TO A POOR CONNECTION. ALL COMPONENTS ON THE HARNESS ASSEMBLY OF THE INFANT WARMER ARE ENCLOSED IN A SHEATH AND FIRE RATED BY (B)(4). SHOULD THE CONNECTORS COMPLETELY FAIL DURING THE OPERATION, AN AUDIBLE AND VISUAL ALARM WILL BE REGISTERED ON THE FRONT CONTROL PANEL OF THE INFANT WARMER, THEREBY ALLOWING THE HOSPITAL STAFF TIME TO ACT AND PROVIDE OTHER MEANS OF WARMING. REPLACEMENT PARTS HAVE SINCE BEEN SENT TO THE MEDICAL CENTER TO REPLACE THE DISCOLOURED HARNESS CONNECTORS.
A MEDICAL CENTER IN (B)(6) REPORTED THAT SIX IW980JEU BABY CONTROL WALL MOUNT INFANT WARMERS HAD DISCOLOURED LOWER AND UPPER HEAD HARNESSES. THIS WAS OBSERVED AFTER PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.
A MEDICAL CENTER IN (B)(6) REPORTED THAT SIX IW980JEU BABY CONTROL WALL MOUNT INFANT WARMERS HAD DISCOLOURED LOWER AND UPPER HEAD HARNESSES. THIS WAS OBSERVED AFTER PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.