ACCESS® 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-00728
- Event Type
- Injury
- Date Received
- March 20, 2011
- Date of Event
- November 28, 2007
- Report Date
- November 29, 2007
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- MMI
- PMA / PMN Number
- K922823/A007
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE SAMPLE WAS COLLECTED IN THE ER INTO BD PLASTIC GEL TUBE. SAMPLE TYPE IS LITHIUM HEPARIN PLASMA. SAMPLE DID NOT APPEAR LIPEMIC OR HEMOLYZED. DRAW VOLUME WAS ADEQUATE. CENTRIFUGATION WAS DONE AT >5, 000 RPM FOR 8 MINUTES. SAMPLE 1 WAS RE-SPUN AT 13, 000 RPM FOR 3 MINUTES PRIOR TO REPEAT ANALYSIS. DILUTION TESTING REPRODUCED VALUES. QC RESULTED IN RANGE PRIOR TO AND FOLLOWING THE EVENT. SYSTEM CHECKS WERE PERFORMED ON (B)(4) 2007, AND RESULTS MET SPECIFICATIONS. CUSTOMER PERFORMED CALIBRATION VERIFICATION FOR THE LOT NUMBER IN USE ON (B)(4) 2007. MATERIAL MET ACCEPTANCE CRITERIA. NO ERRORS WERE RECEIVED IN THE EVENT LOG NEAR THE TIMES OF SAMPLE COMPLETION. ALTHOUGH SERVICE MAID REPAIRS TO THE SYSTEM ON (B)(4) 2007, A CUSTOMER PRODUCT LINE SUPPORT (CPLS) RECEIVED AN ALIQUOT OF THE SAMPLE IN QUESTION AND FOUND THE ADDITION OF BLOCKERS SIGNIFICANTLY REDUCED THE SAMPLE VALUE TO 0.00NG/ML. HETEROPHILE INTERFERENCE HAS BEEN CONFIRMED AS THE ROOT CAUSE OF THIS EVENT. NOTE: THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS FOR ADDITIONAL REPORTABLE EVENTS/OCCURRENCES.
A CUSTOMER CONTACTED BECKMAN COULTER INC., (BCI) REGARDING DISCORDANT TROPONIN (ACCUTNI) RESULTS GENERATED BY THE ACCESS 2 IMMUNOASSAY SYSTEM FOR ONE PATIENT. THE RESULTS WERE REPRODUCIBLE BY THE ACCESS 2 ANALYZER. SAMPLES FROM THE PATIENT WERE TESTED FOR TROPONIN VIA I-STAT AND WERE NEGATIVE. THE PATIENT WAS ADMITTED TO THE ER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS® 2 IMMUNOASSAY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | MMI | BECKMAN COULTER INC. | ACCESS® 2 IMMUNOASSAY SYSTEM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Hospitalization |