FDA Adverse Event Malfunction Summary report: N

AUTOSOFT 90

MDR report key: 20245008 · Received September 18, 2024

Report

Report Number
3003442380-2024-25713
Event Type
Malfunction
Date Received
September 18, 2024
Date of Event
June 17, 2024
Report Date
August 20, 2024
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244018129
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR 1973730- MDR 3003442380-2024-25713- DEVICE 4 OF 10.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN SPAIN. IT WAS REPORTED THAT THE PATIENT EXPERIENCED CANNULA ISSUES WITH TEN INFUSION SETS. THE CANNULAS WERE KINKED. THE EVENT OCCURRED ON 17-JUN-2024, 20-JUN-2024, 29-JUN-2024, 02-JUL-2024, 05-JUL-204, 17-JUL-2024, 26-JUL-2024, AND 08-AUG-2024. PATIENT NOTICED SYMPTOMS WITHIN 3 HOURS OF INSERTION. INFUSIONS SETS WERE USED FOR A FEW HOURS. PATIENT REPLACED THE INFUSION SETS AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. COMPANY DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1319585 AUTOSOFT 90 UNO INSET II 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL A/S 1002817 6004244 05705244018129

Patients

Seq Age Sex Outcome Treatment
1 NA Female