HOMECHOICE
Report
- Report Number
- 1423500-2011-03412
- Event Type
- Malfunction
- Date Received
- March 18, 2011
- Date of Event
- February 22, 2011
- Report Date
- February 22, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). CORRECTED INFORMATION BELOW: DURING EVALUATION OF THE HOMECHOICE (HC) DEVICE, THE PRODUCT ANALYSIS LABORATORY DETERMINED THE HC MACHINE SYSTEM FAILED RETURNED INSTRUMENT TEST/EVALUATION TESTING DUE TO THE FOLLOWING FAILURE: RITE - HEATER BAG TEMPERATURE FAILED FUNCTIONAL SPECIFICATION - TREATMENT STOPPED FOR FLUID TEMPERATURE. NO ALLEGATIONS WERE MADE AGAINST ANY OF THE PATIENT'S DIALYSIS PRODUCTS. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED. THIS FAILURE IS NOT LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, HOWEVER THE EVALUATION HAS NOT BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING EVALUATION OF THE HOMECHOICE (HC) DEVICE, THE PRODUCT ANALYSIS LABORATORY DETERMINED THE HC MACHINE SYSTEM FAILED RETURNED INSTRUMENT TEST/EVALUATION TESTING DUE TO THE FOLLOWING FAILURE: HEATER BAG TEMPERATURE FAILED PERFORMANCE SPECIFICATION AS THE FLUID DELIVERED WAS OUT OF RANGE. THE FAILURE OCCURRED AT 29.9 DEGREES CELSIUS. NO ALLEGATIONS WERE MADE AGAINST ANY OF THE PATIENT'S DIALYSIS PRODUCTS. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |