FDA Adverse Event Malfunction Summary report: N

FREESTYLE LITE

MDR report key: 2024158 · Received March 18, 2011

Report

Report Number
2954323-2011-02363
Event Type
Malfunction
Date Received
March 18, 2011
Date of Event
March 1, 2011
Report Date
April 26, 2011
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED.

Additional Manufacturer Narrative · 1

AN INVESTIGATION WAS PERFORMED ON RETURNED PRODUCT; METER SERIAL NUMBER (B)(4) AND TEST STRIPS LOT NUMBER 1072318. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL FUNCTIONAL TEST RESULTS WERE WITHIN RANGE SPECIFICATION. NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS THE CUSTOMER REPORTED WERE FOUND IN METER'S MEMORY. THIS IS A FINAL REPORT.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 20 MG/DL AND 99 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 1072318