FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE

MDR report key: 20241230 · Received September 17, 2024

Report

Report Number
9610773-2024-32595
Event Type
Malfunction
Date Received
September 17, 2024
Report Date
August 21, 2025
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
PMA / PMN Number
K190744
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

E2/E3: NO INFORMATION AVAILABLE FOR THE OCCUPATION OF THE INITIAL REPORTER OR WHETHER THEY ARE A HEALTHCARE PROFESSIONAL. THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS NOT CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: CHIPPED IN THE LIGHT GUIDE [FIBER] BUNDLE OF DISTAL END AND INSERTION TUBE IS DENT. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, SINCE REPORTED FAILURE WAS NOT CONFIRMED, THE MOST PROBABLE CAUSE OF THE REPORTABLE MALFUNCTION IS NO PROBLEM DETECTED. AND THE MOST PROBABLE CAUSE OF THE MALFUNCTION FOUND DURING DEVICE EVALUATION IS TRACED TO COMPONENT FAILURE. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SUBJECT DEVICE HAD A BENT LIGHT GUIDE, AND IT WAS DARK. THERE WERE NO REPORTS OF PATIENT HARM.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1740657 VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE RIGID VIDEO LAPAROSCOPE HET OLYMPUS WINTER & IBE GMBH WA50052A

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown