FDA Adverse Event Malfunction Summary report: N

INRATIO2

MDR report key: 2023995 · Received February 14, 2011

Report

Report Number
2027969-2011-00319
Event Type
Malfunction
Date Received
February 14, 2011
Date of Event
January 25, 2011
Report Date
February 14, 2011
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
GJS
PMA / PMN Number
K072727
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED. THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND REFERENCE VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. CONCLUSION: ANALYSIS OF THE CLIENT'S DATA FROM INRATIO AND REFERENCE TESTS REVEALED THAT TEST RESULT COMPARISON MET ACCURACY CRITERIA. CUSTOMER'S RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. AS REVIEWED ON 02/12/2011, 32 DISCREPANT RESULT COMPLAINTS WERE REPORTED FOR LOT #232170 YIELDING A COMPLAINT RATE OF 0.016%. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (>0.07%) NO FURTHER ACTION IS REQUIRED AT THIS TIME. THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. CORRECTIVE ACTION NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT RESULTS USING THE INRATIO2 METER: RESULTS AS FOLLOWS: DATE: (B)(6) 2011, PT: 1, INRATIO: 2.4, LAB: 3.3. DATE: (B)(6) 2011, PT: 2, INRATIO: 3.5, LAB: 2.7. DATE: (B)(6) 2011, PT: 3, INRATIO: 2.1, LAB: 4.1. NO PT INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO2 PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC. 100139 232170

Patients

Seq Age Sex Outcome Treatment
1