INRATIO2
Report
- Report Number
- 2027969-2011-00319
- Event Type
- Malfunction
- Date Received
- February 14, 2011
- Date of Event
- January 25, 2011
- Report Date
- February 14, 2011
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- K072727
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED. THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND REFERENCE VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. CONCLUSION: ANALYSIS OF THE CLIENT'S DATA FROM INRATIO AND REFERENCE TESTS REVEALED THAT TEST RESULT COMPARISON MET ACCURACY CRITERIA. CUSTOMER'S RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. AS REVIEWED ON 02/12/2011, 32 DISCREPANT RESULT COMPLAINTS WERE REPORTED FOR LOT #232170 YIELDING A COMPLAINT RATE OF 0.016%. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (>0.07%) NO FURTHER ACTION IS REQUIRED AT THIS TIME. THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. CORRECTIVE ACTION NOT REQUIRED AT THIS TIME.
CALLER ALLEGED DISCREPANT RESULTS USING THE INRATIO2 METER: RESULTS AS FOLLOWS: DATE: (B)(6) 2011, PT: 1, INRATIO: 2.4, LAB: 3.3. DATE: (B)(6) 2011, PT: 2, INRATIO: 3.5, LAB: 2.7. DATE: (B)(6) 2011, PT: 3, INRATIO: 2.1, LAB: 4.1. NO PT INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO2 | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC. | 100139 | 232170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |