FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2023966 · Received March 18, 2011

Report

Report Number
2050012-2011-00755
Event Type
Malfunction
Date Received
March 18, 2011
Date of Event
February 17, 2011
Report Date
February 17, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER STATED THE MAIN VACUUM ERRORS HAVE BEEN ONGOING FOR SEVERAL DAYS AND CAUSE THE INSTRUMENT TO STOP RUNNING. CUSTOMER TECHNICAL SUPPORT (CTS) ASSISTED THE CUSTOMER WITH TROUBLESHOOTING AND RECOMMENDED THE USE OF PERSONAL PROTECTIVE EQUIPMENT BEFORE TROUBLESHOOTING. CTS ASKED THE CUSTOMER TO INSPECT FOR LEAKS. THE CUSTOMER OPENED THE HYDRO AND SAW THAT THE NO FOAM CANISTER LID PLUG IS BUBBLING. CTS ADVISED THE CUSTOMER TO STOP THE INSTRUMENT AND THEN REMOVE AND CLEAN THE NO FOAM PLUG. THE CUSTOMER STATED THIS NO FOAM CANISTER WAS RECENTLY REPLACED AND THAT THE THREADS LOOK OK. THE BUBBLES IN NO FOAM CONTINUE TO FORM AFTER CLEANING. A BCI FIELD SERVICE ENGINEER (FSE) REBUILT PUMPS.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT NO FOAM BUBBLING FROM THE BOTTLE CAP AND LEAKING. THE LEAKING HARDWARE MAY CAUSE INCORRECT RESULTS OR OPERATOR EXPOSURE TO CHEMICALS OR BIOHAZARDS UPON REOCCURRENCE. NO INJURY OR EXPOSURE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER, JJE BECKMAN COULTER INC. DXC600P N/A

Patients

Seq Age Sex Outcome Treatment
1