FDA Adverse Event Malfunction Summary report: N

FLOWART NEEDLE-FREE SPLIT SEPTUM VALVE

MDR report key: 20238412 · Received September 17, 2024

Report

Report Number
3015085713-2024-00006
Event Type
Malfunction
Date Received
September 17, 2024
Date of Event
May 2, 2024
Report Date
September 16, 2024
Manufacturer
ASSET MEDIKAL TASARIM SAN. TIC. A.S.
Product Code
FPA
UDI-DI
08699443580044
PMA / PMN Number
K222889
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE IS NOT AVAILABLE FOR INVESTIGATION AS THE USER HAS DISCARDED IT. WITHOUT THE RETURN OF THE DEVICE, A PROBABLE CAUSE IS UNABLE TO BE DETERMINED. THE DEVICE HISTORY RECORDS DID NOT SHOW ANY CONCERNS REGARDING THE DISCONNECTION. PRODUCT WAS DESIGNED & TESTED ACC. TO ISO 80369-7.

Description of Event or Problem · 0

SIX (6) SEPARATE EVENTS HAVE OCCURRED SINCE MAY 2ND, 2024 WHERE THE DIALYSIS BLOOD LINE BECOMES DISCONNECTED FROM THE FLOWART NEEDLE-FREE SPLIT SEPTUM VALVE WHILE THE PATIENT IS RECEIVING A DIALYSIS TREATMENT. THIS HAS RESULTED IN THE LOSS OF BLOOD ON SEVERAL OF THESE OCCASIONS. THE DISTRIBUTER/MANUFACTURER HAS BEEN NOTIFIED OF EACH EVENT. REFERENCE REPORTS: MW5158285, MW5158286, MW5158287, MW5158288, MW5158289. DURING A ROUTINE DIALYSIS TREATMENT, A NURSE WAS SWABBING THE A1010H-S VALVE WITH ALCOHOL, PER PROCEDURE, WHEN THE VALVE BECAME DISCONNECTED FROM THE BLOOD LINE. THE BLOOD LINE WAS NOT INTENDED TO BE DISCONNECTED AND THE VALVE WAS NOT MANIPULATED IN A MANNER THAT WOULD OTHERWISE RESULT IN A DISCONNECTION. THE A1010H-S VALVE SEEMED TO NOT HAVE BEEN THREADED ONTO THE BLOOD LINE CORRECTLY. THE NURSE WAS EXPERIENCED AND HAS PERFORMED THIS ACTION MANY TIMES WITHOUT A SUDDEN DISCONNECTION OF THE VALVE FROM THE BLOOD LINE. STAFF AT THIS DIALYSIS CENTER HAVE REPORTED THAT THE THREADS OF THE VALVE SOMETIMES CAN BE VERY DIFFICULT TO CONNECT TO SYRINGES OR BLOOD LINES. THEY HAVE NOT REPORTED ISSUES CONNECTING THE VALVE TO THE PATIENT DIALYSIS CATHETER. BASED ON THE DIFFICULTY OF THREADING THE VALVE TO SYRINGES OR BLOOD LINES, THE NURSE BELIEVES THAT THE VALVE DID NOT CONNECT PROPERLY TO THE THREADS OF THE BLOOD LINE, WHICH RESULTED IN THE DISCONNECTION. THE DIFFICULTY IN VALVE ATTACHMENT WAS NOT A MAJOR CAUSE FOR CONCERN UNTIL THE VALVE BECAME DISCONNECTED FOR THE BLOOD LINE. THE PRODUCT WAS NOT RETAINED FOR RETURN TO THE MANUFACTURER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2505816 FLOWART NEEDLE-FREE SPLIT SEPTUM VALVE SET, ADMINISTRATION, INTRAVASCULAR FPA ASSET MEDIKAL TASARIM SAN. TIC. A.S. A1010H-S Z20231211027 08699443580044

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention