FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2023771 · Received March 18, 2011

Report

Report Number
1423500-2011-03391
Event Type
Malfunction
Date Received
March 18, 2011
Date of Event
January 18, 2011
Report Date
February 24, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE HOMECHOICE (HC) DEVICE WAS RECEIVED OPERATIONAL AND IN GOOD CONDITION FOR EVALUATION AT THE PRODUCT ANALYSIS LAB (PAL). A REVIEW OF THE DEVICE LOGS CONFIRMED THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. THE CAUSE FOR THE IIPV FOUND IN THE LOGS WAS DETERMINED TO BE INSUFFICIENT DRAIN DUE TO ONE OR MORE CYCLES ADVANCING TO THE NEXT FILL WHEN SLOW/ NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. THE DEVICE'S SERVICE HISTORY RECORD WAS REVIEWED AND NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RTB-CAPA-(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MEDWATCH WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

TWO INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION WERE IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE (HC) DEVICE. THIS IS REPORT 2 OF 2. THE IIPV OCCURRED ON (B)(6) 2011 DURING DRAIN CYCLE 3. THE PROGRAMMED FILL VOLUME WAS 2000ML AND THE DRAIN VOLUME WAS 3513ML. THIS DRAIN VOLUME MEETS BAXTER?S IIPV CRITERIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 67 YR