FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT EVEROLIMUS-ELUTING CORONARY STENT SYSTEM

MDR report key: 2023766 · Received March 18, 2011

Report

Report Number
2134265-2011-00843
Event Type
Malfunction
Date Received
March 18, 2011
Date of Event
February 22, 2011
Report Date
February 22, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
SIMILAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GB
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: A VISUAL EXAMINATION OF THE DEVICE FOUND THAT THE STENT WAS RETURNED DISLODGED FROM THE STENT DELIVERY SYSTEM (SDS). THE STENT WAS NOT RETURNED. THE BALLOON WAS FOUND TO HAVE THE STENT IMPRESSIONS ON IT AND PILLOWING OF THE BALLOON INDICATING THAT THE STENT WAS CRIMPED PER PROCESS IN THE CORRECT LOCATION DURING MANUFACTURING. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. A PRODUCT MANDREL WAS INSERTED THROUGH THE LUMEN WITH NO RESTRICTIONS NOTED. SOLIDIFIED CONTRAST MEDIA WAS PRESENT WITHIN THE ENTIRE LENGTH OF THE INFLATION LUMEN, THEREFORE INDICATING THE DEVICE HAD BEEN PREPPED FOR USE. THERE WAS A KINK IN THE MIDSHAFT JUST DISTAL TO THE HYPOTUBE SHAFT. THE TIP SECTION OF THE DEVICE WAS VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH IT'S PROFILE THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED HANDLING DAMAGE AS THE EVENT OCCURRED PRIOR TO PATIENT CONTACT. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR A STENTING TREATMENT PROCEDURE A STENT DISLODGEMENT OCCURRED. WHILE PREPPING THE 2.25X28MM PROMUS ELEMENT STENT DELIVERY SYSTEM (SDS) IT WAS NOTED THAT THE STENT WAS NOT ON THE BALLOON. THE STENT WAS UNABLE TO BE LOCATED AND THE DEVICE NEVER ENTERED THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DIFFERENT DEVICE WITH NO PATIENT COMPLICATIONS REPORTED. THE PATIENT'S CURRENT CONDITION IS STABLE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR A STENTING TREATMENT PROCEDURE A STENT DISLODGEMENT OCCURRED. WHILE PREPPING THE 2.25X28MM PROMUS ELEMENT STENT DELIVERY SYSTEM (SDS) IT WAS NOTED THAT THE STENT WAS NOT ON THE BALLOON. THE STENT WAS UNABLE TO BE LOCATED AND THE DEVICE NEVER ENTERED THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DIFFERENT DEVICE WITH NO PATIENT COMPLICATIONS REPORTED. THE PATIENT'S CURRENT CONDITION IS STABLE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROMUS ELEMENT EVEROLIMUS-ELUTING CORONARY STENT SYSTEM STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493911328220 0013305011

Patients

Seq Age Sex Outcome Treatment
1