FDA Adverse Event Malfunction Summary report: N

RECIPROC BLUE FILES, 6X, STERILE

MDR report key: 20235503 · Received September 17, 2024

Report

Report Number
9611053-2024-00173
Event Type
Malfunction
Date Received
September 17, 2024
Report Date
October 11, 2024
Manufacturer
VDW GMBH
Product Code
EKS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE RETURNED RECIPROC BLUE FILE R25 8/100 25MM 025 IS BROKEN AT THE BASE OF THE ACTIVE PART (FATIGUE). NO MATERIAL DEFECT WAS FOUND DURING ANALYSIS OF THE RUPTURE PATTERN. NO UNUSED FILE IS AVAILABLE FOR EVALUATION. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHR REVIEW (BATCH #1871426). ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND. . FOR INFORMATION, WE REMIND THE PRACTITIONER HAS TO MAKE SURE THAT A STRAIGHT-LINE ACCESS IS CREATED PRIOR USING THE RECIPROC BLUE FILES (AS MENTIONED IN THE DFU). ROOT CAUSES ARE NOT IDENTIFIED. ALL COMPLAINTS ARE MONITORED THROUGH THE MONTHLY PRODUCT SURVEILLANCE COMMITTEE AND A CORRECTIVE ACTION COULD BE DETERMINED BY THE COMMITTEE.

Additional Manufacturer Narrative · 0

AS A RESULT OF THIS MALFUNCTION, THE POTENTIAL FOR SURGICAL INTERVENTION EXISTS TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION AS EVIDENCED BY PREVIOUS REPORTED EVENTS WITH SIMILAR FILES. THIS EVENT, THEREFORE, IS REPORTABLE PER 21CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE. ADDITIONAL INFORMATION REGARDING THE PATIENT OUTCOME HAS BEEN REQUESTED AND WILL BE SUBMITTED AS IT BECOMES AVAILABLE.

Description of Event or Problem · 0

IN THIS EVENT IT IS REPORTED THAT A RECIPROC BLUE FILES, 6X, STERILE BROKE DURING USE. THE BROKEN PART REMAINS IN THE CANAL. FURTHER TREATMENT IS PLANNED FOR THE REMOVAL OF THE BROKEN PART. REQUESTED FURTHER INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1831922 RECIPROC BLUE FILES, 6X, STERILE FILE, PULP CANAL, ENDODONTIC EKS VDW GMBH 1871426

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown