FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2023498 · Received March 1, 2011

Report

Report Number
1720753-2011-01801
Event Type
Malfunction
Date Received
March 1, 2011
Date of Event
February 4, 2011
Report Date
March 1, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE GE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE BACKPLANE WAS REPLACED. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM DISPLAYED A MILLIAMPS ON IN ERROR MESSAGE. NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1