FDA Adverse Event Malfunction Summary report: N

GREENLIGHT ADDSTAT

MDR report key: 2023450 · Received March 8, 2011

Report

Report Number
2937094-2011-00338
Event Type
Malfunction
Date Received
March 8, 2011
Date of Event
July 27, 2010
Report Date
August 18, 2010
Manufacturer
AMERICAN MEDICAL SYSTEMS, INNOVATION CENTER - SILICON VALLEY
Product Code
GEX
PMA / PMN Number
K062719
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE EVENT DESCRIPTION, THE CUSTOMER REPORTED TO AMS DID NOT INDICATE A POTENTIAL PT SAFETY ISSUE. THE DEVICE WAS SUBSEQUENTLY RETURNED TO AMS AND ANALYZED. THE INFO FROM THIS FAILURE ANALYSIS WAS RECEIVED ON (B)(6) 2011, AND THE RESULTS OF THE FAILURE ANALYSIS INDICATED THAT AN MDR WAS REQUIRED. THE FAILURE ANALYSIS DISCLOSED THAT THE FIBER CAP REGION BURNED AND/OR MELTED, THE CAP REMAINED INTACT AND ATTACHED. THE SHRINK TUBE AND/OR FIBER JACKET MELTED AND/OR BURNED. THE BEVELED SECTION MELTED AND EXHIBITED BURNT GLUE. THE FIBER CAP GENERALLY EXHIBITED SOME/ALL OF CHAR, DEVITRIFICATION, CRATER AND MELTED GLASS. THIS DEVICE FAILURE IS ASSOCIATED WITH A LACK OF COOLING. THE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE HEAT ACCUMULATION. THE PRODUCT LABELING (PRODUCT INSERT 0127-1410) WARNS THAT TISSUE CONTACT OR TISSUE PROBING MAY CAUSE FIBER DAMAGE OR BREAKAGE. THE PRODUCT LABELING ALSO WARNS OF POSSIBLE CAP DETACHMENT IN THE PT AND INSTRUCTIONS FOR RETRIEVING.

Description of Event or Problem · 1

IT WAS REPORTED BY A CUSTOMER ON (B)(6) 2010, THERE WAS A FLASH OF LIGHT AND THEN THE FIBER DID NOT WORK; THE FIBER HAD NO LASER POWER AT 78,089 JOULES. A FAILURE ANALYSIS, CONDUCTED BY AN AMERICAN MEDICAL SYSTEMS QUALITY ENGINEER, DISCLOSED THAT THE FIBER CAP REGION BURNED AND/OR MELTED, THE CAP REMAINED INTACT AND ATTACHED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GREENLIGHT ADDSTAT SURGICAL FIBER GEX AMERICAN MEDICAL SYSTEMS, INNOVATION CENTER - SILICON VALLEY NA 011H

Patients

Seq Age Sex Outcome Treatment
1 Other