FDA Adverse Event Injury Summary report: N

UNKNOWN ZIMMER HIP

MDR report key: 2023282 · Received February 15, 2011

Report

Report Number
1822565-2011-00616
Event Type
Injury
Date Received
February 15, 2011
Report Date
February 16, 2011
Manufacturer
ZIMMER, INC.
Product Code
JDI
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

INFORMATION WAS RECEIVED FROM A CONSUMER WHO IS NOT REQUIRED TO COMPLETE FORM 3500A. EVALUATION SUMMARY: THE SURGICAL NOTES WERE REVIEWED WHICH DID NOT REVEAL ANY SURGICAL COMPLICATIONS. THE SURGICAL NOTES STATE THAT THE PT HAD A RIGHT BIPOLAR HEMIARTHROPLASTY USING A ZIMMER NATURAL HIP STEM, RIGHT SIZE 1, 48MM BIPOLAR SHELL, 28MM HEAD, MINUS 3.5 NECK LENGTH. X-RAYS WERE NOT PROVIDED; IT IS UNK WHETHER THE COMPONENTS WERE IMPLANTED WITH THE CORRECT FIT AND ORIENTATION AS PER THE SURGICAL TECHNIQUE. PT FACTORS THAT MAY AFFECT THE PERFORMANCE OF THE COMPONENTS SUCH AS AGE, BONE QUALITY, HEIGHT/WEIGHT, ACTIVITY LEVEL, TYPE OF ACTIVITY (LOW IMPACT VS. HIGH IMPACT) ARE UNK. THERE IS INSUFFICIENT INFORMATION TO PERFORM A PROBABLE CAUSE ANALYSIS. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED WITH THE INFORMATION PROVIDED. EVALUATION: REVIEW OF THE DEVICE HISTORY RECORDS WAS NOT POSSIBLE, AS THE LOT NUMBERS WERE NOT PROVIDED. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC. CONSIDERS THE INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT IS REPORTED THAT THE PT IS PRESENTING WITH PAIN AND POSSIBLE INFECTION IN RIGHT HIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNKNOWN ZIMMER HIP HIP PROSTHESIS JDI ZIMMER, INC.

Patients

Seq Age Sex Outcome Treatment
1 Other