FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 2023192
·
Received February 15, 2011
Report
- Report Number
- 1824206-2011-00877
- Event Type
- Malfunction
- Date Received
- February 15, 2011
- Date of Event
- January 24, 2011
- Report Date
- January 24, 2011
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THE TREATMENT FOOT SURFACE CONTROL MODULE (TFSCM) CABLE WAS SEVERED WHICH POSSIBLY DAMAGED THE POWER CONTROL MODULE. THE TECHNICIAN REPLACED THE TFSCM CABLE AND POWER CONTROL MODULE TO REPAIR THE BED.
Description of Event or Problem · 1
THE ACCOUNT ALLEGED THE BED IS BROKEN. THERE IS NO POWER TO THE SIDERAIL FUNCTIONS. NO MOTOR OR BLOWER MOTOR FUNCTIONS AND THE BED IS STUCK IN THE DOWN POSITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |