FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 2023175
·
Received February 15, 2011
Report
- Report Number
- 1824206-2011-00881
- Event Type
- Malfunction
- Date Received
- February 15, 2011
- Date of Event
- February 1, 2011
- Report Date
- February 1, 2011
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN FOUND THE EMPTY BED ON 3RD FLOOR HALLWAY WITH (BROKEN) SIGN ON IT. TOOK THE BED TO BED SHOP. FOUND HI-LO FOOT WOULD DRIFT DOWN. REPLACED THE HI-LO-FOOT UP VALVE TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
FOUND THE EMPTY BED ON 3RD FLOOR HALLWAY WITH "BROKEN" SIGN ON IT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | P1900K006837 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |