FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2023175 · Received February 15, 2011

Report

Report Number
1824206-2011-00881
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
February 1, 2011
Report Date
February 1, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND THE EMPTY BED ON 3RD FLOOR HALLWAY WITH (BROKEN) SIGN ON IT. TOOK THE BED TO BED SHOP. FOUND HI-LO FOOT WOULD DRIFT DOWN. REPLACED THE HI-LO-FOOT UP VALVE TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

FOUND THE EMPTY BED ON 3RD FLOOR HALLWAY WITH "BROKEN" SIGN ON IT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. P1900K006837

Patients

Seq Age Sex Outcome Treatment
1