FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 2023158 · Received February 15, 2011

Report

Report Number
1824206-2011-00876
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
January 24, 2011
Report Date
January 24, 2011
Manufacturer
HILL-ROM INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THAT THE CASTER TEETH WERE WORN. HE REPLACED THE BRAKE/STEER CASTER TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES WHEN THE BRAKES WERE ENGAGED, THE BRAKE/STEER CASTER WOULD STILL SWIVEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY BED BIRTHING BED HDD HILL-ROM INC. 3600

Patients

Seq Age Sex Outcome Treatment
1