FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2023126 · Received February 15, 2011

Report

Report Number
3004209178-2011-80419
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
February 9, 2011
Report Date
February 10, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP STUCK ON REWIND PRIME LOOP. TROUBLESHOOTING WAS PERFORMED. REVIEWED THE ALARM HISTORY AND FOUND SEVERAL ALARMS. RAN A DISPLACEMENT TEST AND THE DEVICE FAILED TWICE. ADVISED THE CUSTOMER TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-522NAP

Patients

Seq Age Sex Outcome Treatment
1 13 YR