HEARTSTART MRX -EMS DEFIBRILLATOR
MDR report key: 2023077
·
Received February 15, 2011
Report
- Report Number
- 1218950-2011-00414
- Event Type
- Malfunction
- Date Received
- February 15, 2011
- Report Date
- January 20, 2011
- Manufacturer
- PHILIPS HEALTHCARE
- Product Code
- MKJ
- PMA / PMN Number
- K031187
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4): IT WAS REPORTED THAT THE CHARGE BUTTON STUCK. THE DEVICE WAS EVALUATED AT PHILIPS. THE SYMPTOM WAS CLARIFIED AS THE DEVICE HANGING WHEN CHARGE BUTTON PRESSED DURING OPCHECK. THE THERAPY PCA WAS REPLACED TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
IT WAS REPORTED THE CHARGE BUTTON STUCK.