FDA Adverse Event Malfunction Summary report: N

HEARTSTART MRX -EMS DEFIBRILLATOR

MDR report key: 2023077 · Received February 15, 2011

Report

Report Number
1218950-2011-00414
Event Type
Malfunction
Date Received
February 15, 2011
Report Date
January 20, 2011
Manufacturer
PHILIPS HEALTHCARE
Product Code
MKJ
PMA / PMN Number
K031187
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): IT WAS REPORTED THAT THE CHARGE BUTTON STUCK. THE DEVICE WAS EVALUATED AT PHILIPS. THE SYMPTOM WAS CLARIFIED AS THE DEVICE HANGING WHEN CHARGE BUTTON PRESSED DURING OPCHECK. THE THERAPY PCA WAS REPLACED TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

IT WAS REPORTED THE CHARGE BUTTON STUCK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTART MRX -EMS DEFIBRILLATOR MKJ PHILIPS HEALTHCARE M3536A

Patients

Seq Age Sex Outcome Treatment
1