FDA Adverse Event Malfunction Summary report: N

ORTHOPEDIC TRAY

MDR report key: 20230413 · Received September 16, 2024

Report

Report Number
3007069947-2024-00002
Event Type
Malfunction
Date Received
September 16, 2024
Date of Event
August 27, 2024
Report Date
November 1, 2024
Manufacturer
AMERICAN CONTRACT SYSTEMS, INC.
Product Code
OJH
UDI-DI
00191072217431
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HISTORY RECORD FOR COMPLAINT LOT WAS REVIEWED, THERE WERE NO ISSUES FOUND. CLEANING LOGS WERE REVIEWED, NO CONTRIBUTING ISSUES WERE DETECTED. LIKELY ROOT CAUSE WAS ASSEMBLY TEAMMATE FAILED TO PROPERLY PERFORM GOWNING SEQUENCE AND HAIR WAS MISSED DURING KIT VERIFICATION. A QUALITY ALERT WAS ISSUED STRESSING IMPORTANCE OF COMPLAINT INCIDENT. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT PRODUCT IS DEFECTIVE OR HAS CAUSED SERIOUS INJURY.

Additional Manufacturer Narrative · 0

THE PRODUCT INVOLVED IN THE EVENT WAS NOT RETURNED FOR EVALUATION. A FOLLOW-UP REPORT WILL BE PROVIDED UPON CONCLUSION OF INVESTIGATION. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT PRODUCT IS DEFECTIVE OR HAS CAUSED SERIOUS INJURY.

Description of Event or Problem · 0

CUSTOMER FOUND FOREIGN MATERIAL IN PACKS WHICH CAUSED A 54 MIN DELAY IN PROCEDURE. DEBRIS WAS DISCOVERED PRIOR TO CASE START, REQUIRING NEW SET UP. NO OTHER PATIENT IMPACT INDICATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2506499 ORTHOPEDIC TRAY ORTHOPEDIC TRAY OJH AMERICAN CONTRACT SYSTEMS, INC. 42-7845311 00191072217431

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown