FDA Adverse Event Malfunction Summary report: N

TERUMO TCM II COOLING AND HEATING SYSTEM

MDR report key: 2022987 · Received February 24, 2011

Report

Report Number
1828100-2011-00428
Event Type
Malfunction
Date Received
February 24, 2011
Date of Event
February 1, 2011
Report Date
February 24, 2011
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
K883603
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

DURING USE OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE USER REPORTED THAT THE HEATER COOLER WOULD NOT COOL. AS A RESULT, ALTERNATE DEVICE WAS EMPLOYED FOR THE PROCEDURE. THE USER REPORTED THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WERE NO ADVERSE CONSEQUENCES TO THE PT AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TERUMO TCM II COOLING AND HEATING SYSTEM HEATING COOLING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 4415

Patients

Seq Age Sex Outcome Treatment
1