FDA Adverse Event Malfunction Summary report: N

ENDEAVOR RESOLUTE RX CORONARY STENT SYSTEM

MDR report key: 2022966 · Received February 24, 2011

Report

Report Number
2953200-2011-00493
Event Type
Malfunction
Date Received
February 24, 2011
Date of Event
January 25, 2011
Report Date
February 3, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
NIQ
PMA / PMN Number
P060033
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL, RESULTS: NO CONCLUSION CAN BE DRAWN. (USE OF FORCE TO ADVANCE THE STENT). (STENT DEFORMATION AND FAILURE TO CROSS THE LESION). CONCLUSION: (DEVICE WAS NOT IDENTIFIED AS THE ROOT CAUSE OF EVENT). (USE OF FORCE TO ADVANCE THE STENT). EVAL SUMMARY: THE DEVICE WAS BENT, WAVY AND KINKED APPROX 16CM DISTAL TO THE STRAIN RELIEF. THE STENT WAS POSITIONED ON THE BALLOON AS PER SPEC. A NUMBER OF STRUTS ON THE 8TH DISTAL STENT SEGMENT WERE RAISED AND DEFORMED. THE 13TH DISTAL STENT SEGMENTS WERE SLIGHTLY RAISED. NO FURTHER DAMAGE NOTED. PLEASE NOTE THAT THIS DEVICE (B)(4) IS DISTRIBUTED OUTSIDE THE UNITED STATES; HOWEVER, IT IS SIMILAR TO THE UNITED STATES DISTRIBUTED PRODUCT (B)(4).

Description of Event or Problem · 1

THE PHYSICIAN WAS ATTEMPTING TO IMPLANT A 2.5 MM DIAMETER X 24MM LENGTH ENDEAVOR RESOLUTE RAPID EXCHANGE (RX) DRUG ELUTING STENT TO TREAT A LESION IN A PT. THE ANGIOGRAPH SHOWED 40% STENOSIS IN THE BODY OF LM, A DIFFUSED LESION OF LAD, RCA AND LCX. USING A GUIDE WIRE, THE DISTAL OF LAD WAS DILATED USING A BALLOON. THE PHYSICIAN ATTEMPTED TO IMPLANT THE ENDEAVOR RESOLUTE RAPID EXCHANGE (RX) DRUG ELUTING STENT THROUGH THE LESION BUT FAILED. RESISTANCE WAS NOTICED WHEN ADVANCING THE DEVICE TO THE TARGET REGION. THE PHYSICIAN USED ANOTHER STENT AND THE OPERATION WAS SUCCESSFUL. NO PT INJURY OCCURRED AND NO CLINICAL SEQUELAE WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDEAVOR RESOLUTE RX CORONARY STENT SYSTEM NIQ MEDTRONIC CARDIOVASCULAR GALWAY NA 0004541356

Patients

Seq Age Sex Outcome Treatment
1 UNK