FDA Adverse Event Malfunction Summary report: N

BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE

MDR report key: 20228788 · Received September 16, 2024

Report

Report Number
9617032-2024-01467
Event Type
Malfunction
Date Received
September 16, 2024
Date of Event
August 19, 2024
Report Date
August 28, 2024
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
50382903643146
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: "MATERIAL #: 364314. LOT/BATCH #: 3122762. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND NO ISSUES WERE OBSERVED RELATING TO CRACKED SYRINGE AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE CRACKED SYRINGE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS."

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING A BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE, A CRACK WAS SEEN IN THE SYRINGE. THE SYRINGE WAS REPLACED TO COMPLETE THE BLOOD COLLECTION. THERE WAS NO REPORT OF ADVERSE IMPACT TO PATIENTS OR USERS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1496819 BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 3122762 50382903643146

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown