FDA Adverse Event Malfunction Summary report: N

COULTER® HMX AUTOLOADER ANALYZER

MDR report key: 2022866 · Received March 17, 2011

Report

Report Number
1061932-2011-00163
Event Type
Malfunction
Date Received
March 17, 2011
Date of Event
February 15, 2011
Report Date
February 15, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKZ
PMA / PMN Number
K922704
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO SAMPLE COLLECTION OR STORAGE INFORMATION WAS PROVIDED. CONTROLS WERE RUN BEFORE AND AFTER THE EVENT, AND WERE WITHIN ACCEPTABLE LIMITS. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(6) 2011 FOR THIS EVENT. THE FSE REPLACED THE PAK LYSE AND PRIMED THE SYSTEM. THE FSE ALSO RAN CONTROLS AND ADJUSTED THE SHEATH FLOW TO OPTIMIZE COUNT TIMES. THE FSE REVISITED AND INDICATED THAT THE MIXING MODULE HAD THREE STICKY ACTUATOR VALVES. THE FSE REPLACED THEM AND CONFIRMED THAT THE ISSUES WERE RESOLVED. ROOT CAUSE IS UNKNOWN. PER LABELING, USE ALL THE FLAGGING OPTIONS (SUSPECT, DEFINITIVE, HIGH/LOW, PARAMETER CODES AND THE LABORATORY'S FLAGGING CRITERIA) TO OPTIMIZE THE SENSITIVITY OF THE INSTRUMENT RESULTS. DO NOT SINGLE OUT ONE MESSAGE OR OUTPUT, SUCH AS A HISTOGRAM OR SCATTERPLOT, TO SUMMARIZE THE SPECIMEN OR PATIENT'S CONDITION.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT THE COULTER® HMX AUTOLOADER ANALYZER OCCASIONALLY GENERATED ERRONEOUS DIFFERENTIAL RESULTS WITH OR WITHOUT INSTRUMENT GENERATED FLAGS. THE RESULTS YIELDED EITHER HIGH MONOCYTES (MO%) OR BASOPHILS (BA%), OR LOW NEUTROPHIL (NE%) RESULTS. MANUAL DIFFERENTIAL RESULTS SHOWED LOWER RESULTS FOR MO% OR BA% AND HIGHER NE% WHEN COMPARED TO INSTRUMENT RESULTS. THERE WERE NO ERRONEOUS RESULTS REPORTED OUT OF THE LABORATORY. THERE WAS NO DEATH, SERIOUS INJURY, OR CHANGE TO PATIENT TREATMENT IN THIS EVENT.