COULTER® HMX AUTOLOADER ANALYZER
Report
- Report Number
- 1061932-2011-00162
- Event Type
- Malfunction
- Date Received
- March 17, 2011
- Date of Event
- February 14, 2011
- Report Date
- February 15, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K922704
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
NO SAMPLE COLLECTION OR STORAGE INFORMATION WAS PROVIDED. CONTROLS WERE RUN BEFORE AND AFTER THE EVENT, AND WERE WITHIN ACCEPTABLE LIMITS. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 FOR THIS EVENT. THE FSE REPLACED THE PAK LYSE AND PRIMED THE SYSTEM. THE FSE ALSO RAN CONTROLS AND ADJUSTED THE SHEATH FLOW TO OPTIMIZE COUNT TIMES. THE FSE REVISITED AND INDICATED THAT THE MIXING MODULE HAD THREE STICKY ACTUATOR VALVES. THE FSE REPLACED THEM AND CONFIRMED THAT THE ISSUES WERE RESOLVED. ROOT CAUSE IS UNKNOWN. PER LABELING, USE ALL THE FLAGGING OPTIONS (SUSPECT, DEFINITIVE, HIGH/LOW, PARAMETER CODES AND THE LABORATORY'S FLAGGING CRITERIA) TO OPTIMIZE THE SENSITIVITY OF THE INSTRUMENT RESULTS. DO NOT SINGLE OUT ONE MESSAGE OR OUTPUT, SUCH AS A HISTOGRAM OR SCATTERPLOT, TO SUMMARIZE THE SPECIMEN OR PATIENT'S CONDITION.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT THE COULTER HMX AUTOLOADER ANALYZER OCCASIONALLY GENERATED ERRONEOUS DIFFERENTIAL RESULTS WITH OR WITHOUT INSTRUMENT GENERATED FLAGS. THE RESULTS YIELDED EITHER HIGH MONOCYTES (MO%) OR BASOPHILS (BA%), OR LOW NEUTROPHIL (NE%) RESULTS. MANUAL DIFFERENTIAL RESULTS SHOWED LOWER RESULTS FOR MO% OR BA% AND HIGHER NE% WHEN COMPARED TO INSTRUMENT RESULTS. THERE WERE NO ERRONEOUS RESULTS REPORTED OUT OF THE LABORATORY. THERE WAS NO DEATH, SERIOUS INJURY, OR CHANGE TO PATIENT TREATMENT IN THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER® HMX AUTOLOADER ANALYZER | AUTOMATED DIFFERENTIAL CELL COUNTER | GKZ | BECKMAN COULTER, INC. | HMX AL 115V | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |