HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-03345
- Event Type
- Malfunction
- Date Received
- March 17, 2011
- Date of Event
- February 23, 2011
- Report Date
- February 23, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR (SE) 2240 (AIR IN SET) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE; HOWEVER, PER THE COMPLAINT INFORMATION THE CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER THE PATIENT DISCONNECTED THE PATIENT LINE FROM THE TRANSFER SET. A LOT NUMBER WAS UNKNOWN THEREFORE A BATCH REVIEW WAS NOT PERFORMED. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). THE SAMPLE WAS DISCARDED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THEN A FOLLOW UP MDR WILL BE SUBMITTED.
A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER TO REQUEST ASSISTANCE ON THE HOME CHOICE (HC). PER THE INITIAL REPORT, THE HOME PATIENT (HP) HAD QUESTIONS ON A SYSTEM ERROR (SE) 2240 IN DWELL 3 OF 4. THE HP DISCONNECTED AND WENT BACK TO BED. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM. HP REMOVED THE CASSETTE AND BAG FROM THE HC. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. THE WRITER CONTACTED THE HOME PATIENT (HP) ON (B)(6) 2011 REGARDING THE REPORTED PROBLEM. THE HP SAID HE KEEPS THE BAGS ALL LAYERED PROPERLY AND MADE SURE THE CONNECTIONS WERE TIGHT. THE HP SAID HE WENT OVER THE SUPPLIES AND EVERYTHING LOOKED PROPER. HE DID NOT KNOW HOW AIR GOT INTO THE LINE. THE HP SAID THE LOT NUMBER WAS UNKNOWN AND NO SAMPLE WAS AVAILABLE. THE HP SAID EVERYTHING WAS WORKING FINE NOW AND IT WAS ALL TAKEN CARE OF. THE NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | PD CYCLER |