GENDEX EXPERT DC INTRAORAL X-RAY SYSTEM
Report
- Report Number
- 3004115000-2011-00006
- Event Type
- Malfunction
- Date Received
- March 17, 2011
- Date of Event
- February 16, 2011
- Report Date
- February 17, 2011
- Manufacturer
- GENDEX DENTAL SYSTEMS
- Product Code
- EHD
- PMA / PMN Number
- K992610
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- DENTIST
Narratives
UPDATED INFORMATION WAS OBTAINED FROM THE DEALER SERVICE TECHNICIAN FROM THE RE-INSTALLATION OF THE UNIT. FOR THE ORIGINAL INSTALLATION THE UNIT WAS NOT MOUNTED WITH THE OPTIONAL 16" WALL PLATE. THE UNIT WAS INSTALLED TO THE WALL WHICH CONTAINED TWO 5/8TH SHEETS OF PLYWOOD, ONE ON TOP OF THE OTHER, SANDWICHED BETWEEN THE 2X4 STUDS. THE UPPER AND LOWER LAG BOLTS WERE SCREWED INTO THE PLYWOOD. THE TOP LAG BOLT PULLED OUT OF THE PLYWOOD CAUSING THE INCIDENT OF THE UNIT FALLING OFF THE WALL DURING USE. THE UNIT WAS REPAIRED AND RE-INSTALLED BY THE DEALER SERVICE TECHNICIAN USING A 16" WALL PLATE MOUNTED TO THE 2X4 STUDS ON 16" CENTER.
THE ASSISTANT BUMPED HER HAND ON THE CHAIR WHEN CATCHING THE UNIT AS IT FELL. DURING A FOLLOW-UP WITH THE OFFICE, IT WAS CONFIRMED THAT THE ASSISTANT'S HAND WAS FINE. NO MEDICAL TREATMENT WAS NEEDED. THE TOP LAG BOLT PULLED OUT OF THE WALL PUTTING THE WEIGHT FORCE OF THE X-RAY UNIT ON THE BOTTOM LAG CAUSING THE CASTING AT THE BOTTOM LAG MOUNT TO BREAK ALLOWING THE UNIT TO FALL. DURING A FOLLOW-UP CALL WITH THE DEALER SERVICE TECHNICIAN, HE STATED THAT THERE WAS NO APPARENT REASON WHY THE TOP LAG PULLED OUT, THEY WERE USING 1.5" BACKING ON A SINGLE STUD MOUNT AND THE LAGS APPEARED TO BE DRILLED IN THE CENTER OF THE 2X4 STUD. LABELING PROVIDED WITH THE X-RAY UNITS INSTRUCTS THE INSTALLERS TO VERIFY THE WALL SUPPORT CAPABILITY PRIOR TO INSTALLATION. IN CONCLUSION, THIS WAS AN INSTALLATION ISSUE AND NOT A MALFUNCTION OF THE X-RAY UNIT. THE X-RAY UNIT WAS REPAIRED AND RE-INSTALLED BY THE DEALER SERVICE TECHNICIAN THE DAY FOLLOWING THE INCIDENT. DEVICE REPAIRED BY DEALER SERVICE TECH.
THE X-RAY UNIT FELL FROM THE WALL DURING USE AND WAS CAUGHT BY THE ASSISTANT WHEN IT FELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENDEX EXPERT DC INTRAORAL X-RAY SYSTEM | DENTAL X-RAY EQUIPMENT | EHD | GENDEX DENTAL SYSTEMS | EXPERT DC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |