FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 2022521 · Received February 15, 2011

Report

Report Number
2027969-2011-00335
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
January 24, 2011
Report Date
February 15, 2011
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
GJS
PMA / PMN Number
021923
Removal / Correction Number
NI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: INRATIO: 3.6, REFERENCE: 2.8, MEAN: 3.20, CONFIDENCE LIMITS: 1.9 - 4.6. ANALYSIS OF CUSTOMER'S DATA REVEALED THAT INRATIO AND REFERENCE TEST RESULT COMPARISON DID MEET ACCURACY CRITERIA. CUSTOMER'S RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. NO PRODUCT IS EXPECTED TO BE RETURNED. NO FURTHER INVESTIGATION REQUIRED. AS REVIEWED ON (B)(6) 2011, EIGHTY-THREE DISCREPANT RESULT COMPLAINTS WERE REPORTED FOR LOT #243934, YIELDING A COMPLAINT RATE OF 0.062%. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA CORRECTIVE ACTION (>0.07%), NO FURTHER ACTION IS REQUIRED AT THIS TIME. THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: (B)(6) 2011, INRATIO: 3.6, LAB: 2.8.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC. 100071 243934

Patients

Seq Age Sex Outcome Treatment
1 NI