Additional Manufacturer Narrative · 1
THE TECH FOUND THE CASTERS WERE NOT HOLDING DUE TO A BROKEN BRAKE/STEER LINKAGE ON THE FOOT END. HE USED A BRAKE/STEER UPGRADE TO REPAIR THE STRETCHER.
Description of Event or Problem · 1
THE ACCOUNT ALLEGED THE CASTERS WILL NOT COME OUT OF BRAKE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRANSTAR STRETCHER | WHEELED STRETCHER | FPO | HILL-ROM, INC. | 8000 |